首页 > 全球网络 > 上海 > 出版刊物 > 专业文章 > 医疗器械企业出海巴西系列文章二:巴西医疗器械风险分类标准与市场准入路径

医疗器械企业出海巴西系列文章二:巴西医疗器械风险分类标准与市场准入路径

作者:瞿颖影 尹洁 2026-11-23

本系列文章一已梳理 ANVISA 的法律地位、巴西医疗器械规范的五层结构,以及 B-GMP 认证、UDI/SIUD、AREE 信赖路径与 INMETRO/ANATEL 平行认证等关键合规节点。本文承接其中的“风险分级与准入路径”一项,就 RDC 751/2022 附件一的二十二条分类规则、第 8 条的分类适用原则,以及备案与注册两条路径的程序规则逐条展开,并以汇总表形式呈现全部分类规则,供中国医疗器械企业在确定产品等级与申报路径时直接对照巴西法条使用。

Article One of this series set out ANVISA’s legal status, the five-tier structure of Brazilian device regulation, and the key compliance checkpoints — B-GMP certification, UDI/SIUD, the AREE reliance pathway, and the parallel INMETRO/ANATEL certifications. This article takes up one item from that overview — risk classification and the access routes — and works through the twenty-two classification rules of Annex I to RDC 751/2022, the classification principles of Article 8, and the procedural rules of the notification and registration routes, presenting the full rule set in a summary table for PRC manufacturers to use directly when fixing a product’s class and filing route.


引言/Introduction


在巴西,产品的风险等级一经确定,其后的全部合规成本即随之锁定:适用备案还是注册、是否需在批准前取得 B-GMP 证书、技术档案的深度、有效期与续展义务、乃至 UDI 赋码与数据传输的时间表,均由等级决定。分类因此不是申报流程的第一步,而是整个巴西项目的定价前提。实务中最常见的延误,并非来自技术文件的缺失,而是来自企业按国内或欧盟的分类结论直接套用于巴西,至审评阶段方被 ANVISA 重新归类,进而触发整套材料的返工。

In Brazil, once a product’s risk class is fixed, the whole downstream compliance cost is fixed with it: whether the product takes the notification or the registration route, whether a B-GMP certificate must precede approval, how deep the technical dossier must go, the validity period and renewal duty, and even the UDI assignment and data-transmission timetable. Classification is therefore not the first step of a filing — it is the pricing premise of the entire Brazilian project. The most common cause of delay in practice is not a missing technical document but a manufacturer importing its domestic or EU classification conclusion wholesale into Brazil, only for ANVISA to re-classify the product during review and send the whole package back for rework.


在法律渊源上,分类制度并非始于 ANVISA 的决议。《第 6.360 号法》(Lei nº 6.360, de 23 de setembro de 1976)第 12 条已确立“未经注册不得工业化、上架销售或交付”的基本规则,第 25 条则规定医疗器械(correlatos)在制造或进口前须先被判定是否应予注册。换言之,“该不该注册、按哪种程序注册”自一九七六年起即是一个法定前置问题,而非行政裁量事项;RDC 751/2022 所做的是为这一判定提供可操作的规则。¹

In terms of legal sources, classification did not begin with ANVISA’s resolutions. Article 12 of Lei nº 6.360, of 23 September 1976 already established the basic rule that an unregistered product may not be industrialized, offered for sale or delivered for consumption, and Article 25 requires medical devices (correlatos) to be assessed, before manufacture or import, as to whether registration is required. "Whether registration is needed, and under which procedure" has therefore been a statutory precondition since 1976 rather than a matter of administrative discretion; what RDC 751/2022 adds is an operable set of rules for making that determination.¹


分类的现行依据集中于 RDC 751/2022 第 6 条至第 9 条及附件一:第 6、7 条划定两条准入路径,第 8 条规定分类规则的适用原则,第 9 条则将具体分类交由附件一的二十二条规则完成。本文即以此四条为轴,并先行交代其适用前提——第 2 条的范围与第 4 条的定义。²

The current basis sits in Articles 6 to 9 of RDC 751/2022 and its Annex I: Articles 6 and 7 set the two access routes, Article 8 states the principles governing how the rules apply, and Article 9 delegates the actual classification to the twenty-two rules of Annex I. These four provisions are the spine of this article; we begin with their precondition — the scope in Article 2 and the definitions in Article 4.²



一、医疗器械的定义与适用范围(RDC 751/2022 第 2、4 条)

I. Definition and Scope of Medical Devices (RDC 751/2022, Arts. 2 and 4)


1.1 适用范围与排除情形(第 2 条)

1.1 Scope and Exclusions (Article 2)


第 2 条界定适用范围:本决议适用于其所定义的医疗器械,且依风险分类,备案或注册为强制性义务。该条进一步以多款排除特定形态——二手或翻新器械(§2,适用 RDC 579/2021)、定制器械(§3,适用 RDC 305/2019)、体外诊断器械(§4,现由 RDC 830/2023 承接)、药品/人体细胞组织血液/化妆品/消毒剂/食品(§5);同时明确,用于美学矫正与美化的有源器械(§6)以及专用于医疗器械清洁、消毒或灭菌的有源器械(§7)亦属医疗器械。³ 对中国企业而言,最易错置的是最后两款:国内常作为美容设备或清洗消毒设备管理、未必按医疗器械申报的产品,在巴西明确落入医疗器械范畴。

Article 2 defines scope: the resolution applies to the medical devices it defines, and notification or registration is mandatory by risk class. It carves out used/reconditioned devices (§2, RDC 579/2021), custom-made devices (§3, still referring to RDC 305/2019, which has been repealed by RDC 925/2024), in vitro diagnostic devices (§4, now carried by RDC 830/2023), and medicines/cells/tissues/blood/cosmetics/sanitizers/food (§5). Active devices for aesthetic correction/beautification (§6) and active devices specifically intended to clean, disinfect or sterilize medical devices (§7) are themselves medical devices.³ For PRC enterprises the last two paragraphs are the easiest to get wrong: products treated domestically as beauty equipment or as cleaning/disinfection equipment, and not necessarily filed as medical devices, fall squarely within the medical-device perimeter in Brazil.


体外诊断器械须特别注意:IVD 已自 RDC 751/2022 中分离,适用 RDC 830/2023 的独立分类规则。本文以下所述二十二条规则均不适用于 IVD。⁴

IVDs need a specific warning: they have been separated out of RDC 751/2022 and are governed by the independent classification rules of RDC 830/2023. None of the twenty-two rules discussed below applies to an IVD.⁴


1.2 关键定义(第 4 条)

1.2 Key Definitions (Article 4)


第 4 条对医疗器械的实质性定义与欧盟 MDR 高度对齐,因此持有 CE 认证的产品通常不会在“是否为医疗器械”这一层面出现分歧,分歧几乎总是发生在下一层面——具体等级。真正与分类直接相关的是“附件”(acessório)的定义:制造商意欲与一件或多件特定医疗器械联合使用、以便或协助该器械依预期用途使用的产品。该定义是第 8 条 §2“附件独立分类”的适用前提:一件在国内作为主机零部件随机申报的物品,如构成巴西法上的附件,则须单独定级,且其等级并不当然低于主机。⁵

The substantive definition of a medical device in Article 4 is closely aligned with the EU MDR, so a CE-marked product rarely generates a dispute at the level of "is this a medical device"; the dispute is almost always at the next level down — which class. The definition that bears directly on classification is accessory (acessório): a product intended by its manufacturer to be used jointly with one or more specific devices to specifically and directly enable or assist their intended use. That definition is the precondition for the "accessories are classified independently" rule in Article 8 §2: an item filed domestically as a component bundled with the main unit must, if it qualifies as an accessory in Brazilian law, be classified on its own — and its class is not necessarily lower than the main unit’s.⁵


二、风险分类标准(RDC 751/2022 第 5–9 条及附件一)

II. Risk Classification Standards (RDC 751/2022, Arts. 5–9 and Annex I)


2.1 分类依据与等级划分

2.1 Basis of Classification and Risk Classes

image.png


2.2 分类规则的适用原则(第 8 条)

2.2 Principles for Applying the Classification Rules (Article 8)


第 8 条⁸规定分类须依医疗器械的“预期用途”(finalidade prevista)适用规则,并对若干边界情形设定原则:若器械拟与另一器械联合使用,分类规则分别适用于每一器械(§1);器械之附件须独立分类,不依附于其所配合之器械(§2);控制或影响某器械使用的软件,分类与该器械相同(§3);独立软件(SaMD)单独分类(§4);非专用于或主要作用于身体特定部位的器械,按最关键的用途分类(§5);同一器械适用多条规则时,取导致更高分类的最严格规则(§6);“连续使用”时长的计算规则(§7:使用中的临时中断或为清洁消毒而临时取下不中断计时;制造商拟以同类器械即时替换的,累计计算),瞬时(<60 分钟)、短期(60 分钟至 30 天)、长期(>30 天)则由第 4 条定义;可独立作出直接诊断的器械之界定(§8)。

Article 8⁸ directs that classification follows the device’s intended purpose, with boundary principles: devices used with another device are classified separately (§1); accessories are classified independently (§2); software controlling or influencing the use of a device shares its class (§3), while standalone software (SaMD) is classified independently (§4); devices not exclusively or primarily aimed at a body part are classified by the most critical use (§5); where several rules apply, the most stringent (higher class) prevails (§6); continuous duration disregards temporary interruptions or removal for cleaning/disinfection, and is aggregated where the device is intended to be replaced immediately by another of the same type (§7), with transient (<60 min), short-term (60 min–30 days) and long-term (>30 days) use defined in Article 4; and devices providing a diagnosis on their own or supplying decisive information for diagnosis are defined (§8).


其中有三条在实务中的杀伤力远超其文字长度。第一,§6 的“最严格规则优先”意味着分类结论取决于最高而非最典型的那一种用法,因此说明书中一句宽泛的适应症描述,可能直接将产品从 II 类推向 IV 类。第二,§7 规定“连续使用”不因临时中断或清洁消毒取下而中断,制造商拟以同类器械即时替换使用的,替换前后时长累计计算,从而跨越 30 天的长期门槛。第三,§2 的附件独立分类意味着一个项目可能需要多份申报而非一份。起草说明书与预期用途声明时,应将这三条作为核验清单逐项比对。

Three of these carry far more practical weight than their length suggests. First, the "most stringent rule prevails" principle in §6 means the classification turns on the highest, not the most typical, use — so a single broadly drafted indication in the instructions for use can push a product straight from Class II to Class IV. Second, under §7 continuous use is not broken by temporary interruptions or removal for cleaning, and use of a device intended to be replaced immediately by another of the same type is aggregated, crossing the 30-day long-term threshold. Third, independent classification of accessories under §2 means a project may require several filings rather than one. These three should be run as a checklist against the draft instructions for use and the intended-purpose statement.


2.3 附件一分类规则汇总表(规则 1–22)

2.3 Summary Table of Annex I Classification Rules (Rules 1–22)


附件一以二十二条规则构成巴西分类体系。需注意其双重渊源:整体框架对接 IMDRF 的监管逻辑,而附件一的规则结构与表述则与欧盟 MDR 附件八高度贴近。这一特征对已持有 CE 认证的企业是个好消息——欧盟的分类论证可作为起点;但不应误认为两者逐条对应,规则编号虽与 MDR 附件八一致,但等级映射与部分分支条件不同(如规则 14、18 在巴西归 IV 类),仍须以附件一葡文原文逐条核对。⁷ 各规则的适用器械类型与风险等级范围见下表。

Annex I contains twenty-two rules. Note their dual provenance: the overall framework follows IMDRF regulatory logic, while the structure and wording of the Annex I rules track EU MDR Annex VIII closely. That is good news for a company already holding CE marking — the EU classification reasoning is a usable starting point — but the two should not be assumed to correspond rule by rule; the numbering matches MDR Annex VIII, but the class mapping and some branch conditions differ (Rules 14 and 18, for example, lead to Class IV in Brazil), and the Portuguese text of Annex I still has to be checked line by line.⁷ Each rule’s device type and class range is set out in the table below.

image.png
image.png


注:“风险等级”栏所示为该规则项下可能的风险等级范围,具体定级取决于器械的预期用途、使用时长与接触部位等因素;同一器械适用多条规则的,依第 8 条 §6 取最高等级。

Note: The "Risk Class" column shows the possible class range under each rule; the actual class depends on intended purpose, duration and contact site. Where several rules apply to one device, Article 8 §6 takes the highest class.


2.4 特殊形态的分类处理

2.4 Classification Treatment of Special Device Forms


软件(SaMD):依规则 11 单独分类;独立的软件医疗器械按自身风险定级(第 8 条 §4)。另有 RDC 657/2022 针对 SaMD 作出专门规定(标签、使用说明、安全与性能要求等)。⁹ 附件:依第 2 条 §1 与第 8 条 §2,附件须独立分类;但制造商专门为其已备案/已注册器械配套制造、且技术档案中已含该附件信息的附件,可豁免单独备案/注册(第 2 条 §10),新附件亦可纳入原备案/注册流程(§11)。³ 定制器械与二手/翻新器械:分别适用 RDC 305/2019 与 RDC 579/2021,不适用 RDC 751/2022 的分类与注册路径(第 2 条 §2–§3)。¹⁰ 家族、系统与成套(kit)器械:ANVISA 亦可对医疗器械家族、系统与成套器械授予备案或注册(第 3 条),具体归组依专门规范。¹¹

Software (SaMD): classified under Rule 11; standalone SaMD is classified on its own merits (Art. 8 §4). RDC 657/2022 provides specific SaMD provisions (labelling, instructions, safety/performance).⁹ Accessories: classified independently (Art. 2 §1; Art. 8 §2). Accessories made exclusively to integrate a manufacturer’s already-notified/registered devices and covered in the technical dossier are exempt from separate notification/registration (Art. 2 §10); new accessories may be added to the original process (§11).³ Custom-made and used/reconditioned devices: governed respectively by RDC 925/2024 (replacing RDC 305/2019) and RDC 579/2021, outside the RDC 751/2022 classification and registration pathway (Art. 2 §2–§3).¹⁰ Families, systems and kits: ANVISA may also grant notification/registration to families, systems and sets (Art. 3), grouped per specific regulation.¹¹


2.5 风险等级的下游效应

2.5 Downstream Effects of the Risk Class


分类之所以是定价前提,在于等级一经确定,下表所列各项义务即随之固定。下表将分散于多部规范的结果按等级归并,供项目排期与预算测算使用。 / Classification is the pricing premise because, once the class is fixed, every obligation in the table below is fixed with it. The table consolidates results scattered across several instruments into a single class-indexed view, for use in project scheduling and budgeting.

image.png

注:“CBPF 不作为批准前提”不等于 I/II 类免于 B-GMP——第 10 条 §7 要求备案的维持同样以遵守 B-GMP 为条件,差别仅在核验时点(事前审评或事后稽查)。

Note: "Not a condition of grant" does not exempt Class I/II from B-GMP— Article 10 §7 conditions the maintenance of a notification on B-GMP compliance as well – or clinical data (RDC 848/2024); only the moment of verification differs (prior review or subsequent inspection).


三、两条准入路径的程序规则(RDC 751/2022 第 10–11 条)

III. Procedural Rules of the Two Access Routes (RDC 751/2022, Arts. 10–11)


系列文章一已给出两条路径的结果性差别(备案免续展、注册 10 年可续展,技术档案是否随申请提交)。以下补充的是程序层面的差别——这部分直接决定项目排期与被退回的风险。

Article One already set out the difference in outcomes between the two routes (notification does not expire; registration runs ten years and is renewable; whether the technical dossier accompanies the application). What follows adds the procedural difference — the part that drives project scheduling and the risk of rejection.


3.1 备案制(Notificação)—— I 类与 II 类

3.1 Notification (Notificação) — Class I and Class II


第 10 条 §5 明确规定:对备案及备案变更申请不作技术审查(não haverá análise técnica),产品在满足要求后即视为已合规化(regularizado);但 ANVISA 得随时对备案流程及其变更进行文件性或税收性评估,并在必要时要求补充信息或澄清。第 10 条 §6 规定:备案的处理在申请人提交后最长 30(三十)日内常规完成。¹³

Article 10 §5 states expressly that there is no technical analysis of notification and notification-alteration applications; the products are deemed regularized once the requirements are met — without prejudice to documentary reviews or regulatory inspections at any time, and ANVISA may demand additional information or clarification. Article 10 §6 sets the routine processing of a notification at up to 30 (thirty) days after submission.¹³


但“不作技术审查”不等于“不必满足实体要求”。第 10 条 §7 规定,备案与注册的维持均以遵守 B-GMP、基本安全与性能要求及专门规范为前提。I/II 类器械的合规义务并未减少,只是核验时点由事前审评移至事后稽查与警戒——而一旦暴露,往往已伴随实际发生的销售与库存。¹³

But "no technical analysis" is not "no substantive requirements." Article 10 §7 ties the maintenance of both notification and registration to compliance with B-GMP, the essential safety and performance requirements, and the specific norms. The burden on Class I/II devices is not reduced; only the moment of verification moves, from prior review to subsequent inspection and vigilance — by which point exposure usually comes with sales and inventory already in place.¹³


3.2 注册制(Registro)—— III 类与 IV 类

3.2 Registration (Registro) — Class III and Class IV


III 类、IV 类器械须走完整的上市授权路径。第 10 条 §1 规定:ANVISA 须对注册、注册变更或再验证(revalidação)的文档进行审评,并以官方途径出具意见。第 10 条 §3 允许 ANVISA 基于技术理由、为证明产品安全与性能、因潜在公共健康风险,要求提交补充文件与信息。具体应提交的文件清单则规定于第 13 条(备案)与第 14 条(注册)(注册须提交 ANVISA 签发的 CBPF 或其申请受理凭证;属 SBAC 强制认证范围的产品,备案与注册均须附 SBAC 合格证书,参见系列文章一)。¹³

Class III/IV devices require the full marketing-authorization path. Article 10 §1 requires ANVISA to assess the documentation for registration, alteration or revalidation of registration and to issue its opinion through official means. Article 10 §3 allows ANVISA, for technical reasons, to prove the product’s safety and performance, and on grounds of potential public-health risk, to demand additional documents and information. The actual document lists sit in Article 13 (notification) and Article 14 (registration) — registration requires the ANVISA CBPF or its filing receipt, and products under compulsory SBAC certification must attach the SBAC conformity certificate on either route, as discussed in Article One.¹³


对排期最具实务意义的是第 10 条 §4:申请若文件缺失、表格或声明填写不全、信息缺失或不可读、文件过时、应附合规证书而未附,或创新技术/适应症产品缺乏临床证据者,ANVISA 不发出技术补正要求(exigência técnica),而直接作出不予同意(não anuência)或简易驳回(indeferimento sumário)。被此条退回的申请得不到任何实质性审评意见,企业既无从获知技术缺陷,又须重新排队。¹³

The provision with the greatest scheduling impact is Article 10 §4: applications with missing documents, incomplete forms or declarations, missing or illegible information, obsolete documents, a missing compliance certificate where one is required, or — for innovative technologies or indications — an absence of clinical evidence, are not eligible for a technical deficiency letter (exigência técnica) and are summarily disapproved or rejected. An application rejected under this paragraph receives no substantive review opinion, so the manufacturer learns nothing about its technical shortcomings and must requeue.¹³


语言方面需补充一点:申请表、使用说明或用户/操作手册及标签样稿须以葡萄牙语提交(第 10 条 §9)为人所熟知,但第 10 条 §10 另规定,前述之外的其他文件可以葡萄牙语、西班牙语或英语提交,具体依专门规范确定。这意味着技术支撑材料(性能验证、生物相容性、临床评价等)的英文版本可直接复用,无须全量译为葡文——这是项目预算中常被高估的一项成本。¹³

One point on language should be added: that petition forms, instructions for use or user manuals and label artwork must be in Portuguese (Article 10 §9) is well known, but Article 10 §10 separately permits the remaining documents to be submitted in Portuguese, Spanish or English, under rules set by specific regulation. Supporting technical material — performance validation, biocompatibility, clinical evaluation — can therefore be reused in English rather than translated in full, which is a cost routinely overestimated in project budgets.¹³

第 11 条规定:注册自其在《联邦官方公报》(DOU)公布之日起有效期为 10(十)年,并得以相同期限连续再验证(revalidação)。¹⁴ 第 10 条 §8 则规定,注册的批准(deferimento)以 ANVISA 签发的 B-GMP 证书之公布为条件。两条合读的含义是:分类结论一旦落在 III/IV 类,工厂端的 B-GMP 准备周期即成为整个项目的关键路径,而非产品注册本身。¹³

Article 11 fixes the registration’s validity at 10 (ten) years from its publication in the Diário Oficial da União (DOU), revalidable successively for equal periods.¹⁴ Article 10 §8 makes the grant of the registration conditional on publication of the B-GMP certificate issued by ANVISA. Read together, the consequence is this: once the classification lands in Class III or IV, the factory-side B-GMP lead time — not the product registration itself — becomes the critical path of the entire project.¹³


3.3 与等级挂钩的实体门槛:RDC 848/2024 的临床证据要求

3.3 The Class-Linked Substantive Threshold: Clinical Evidence Requirements under RDC 848/2024


系列文章一已说明 RDC 848/2024 是现行的安全与性能实体标准。此处只补充与分类直接相关的一点:对 III 类与 IV 类器械,符合性须以临床数据支持——或为对现有文献的批判性评估,或为针对该产品开展的临床试验。这一要求与第 10 条 §4 相互咬合,共同构成 III/IV 类产品的真实门槛。¹⁵

Article One already identified RDC 848/2024 as the governing substantive safety-and-performance standard. Only the point tied directly to class needs adding here: under Articles 15–16 of RDC 848/2024, clinical data are required for medical devices generally, including Classes III and IV, supported by one or more clinical evaluation reports critically analysing relevant scientific literature and/or clinical investigation reports. Device-specific investigations are required in the circumstances set out in Articles 18 and 19, including the Class III/IV cases specified in Article 19, item II. Read with Article 10 §4, this forms the real threshold for Class III/IV products.¹⁵


因此应在项目起步阶段做一项判断:拟在巴西主张的适应症,是否已被国内或其他法域的临床资料涵盖。若拟申适应症宽于现有临床证据所支持的范围,则或需限缩预期用途声明(这反过来可能降低分类等级),或需补做临床。两者成本差异往往以年计,故分类、预期用途与临床证据三者应在同一份决策文件中一并考虑,而不宜分别交由注册、临床与市场部门各自决定。

One judgement therefore belongs at project inception: is the indication to be claimed in Brazil already covered by the clinical record generated domestically or in other jurisdictions? If the Brazilian claim is broader than the existing evidence supports, the choice is either to narrow the intended-purpose statement (which may in turn lower the class) or to run additional clinical work. The cost difference is typically measured in years, which is why classification, intended purpose and clinical evidence belong in a single decision document rather than being settled separately by the regulatory, clinical and marketing teams.


四、结语:分类错误的代价与实务要点

IV. Conclusion: The Cost of Misclassification and Practical Points

分类错误在巴西不是一个可以在审评中“顺手更正”的问题。如果企业按 I/II 类完成备案并已实际销售,而 ANVISA 在后续稽查中认定其应属 III/IV 类,则该产品自始至终处于未获授权状态。这意味着已发生的进口与销售构成 Lei nº 6.437/1977 项下的卫生违法,可能导致罚款、产品扣押与销毁、暂停商业化,并须自零启动注册程序——而注册又以 B-GMP 证书的公布为前提。换言之,分类错误的真实代价不是重新填表,而是数年的市场空窗加上已售批次的历史责任。¹⁶

Misclassification is not something that can be quietly corrected mid-review in Brazil. If a manufacturer has completed a Class I/II notification and begun selling, and ANVISA subsequently determines on inspection that the product belongs in Class III/IV, the product has been unauthorized throughout. Imports and sales already made constitute a sanitary infraction under Lei nº 6.437/1977, exposing the company to fines, seizure and destruction of product, and suspension of commercialization — and the registration process must then start from zero, itself conditional on publication of a B-GMP certificate. The true cost of misclassification is therefore not a refiled form but a market gap measured in years, plus historical liability for batches already sold.¹⁶


由此归纳三项实务要点:其一,分类应在商业决策阶段完成,而非在申报准备阶段完成,因为等级决定了 B-GMP、临床证据与时间表三项核心成本。其二,预期用途声明应按巴西申报需要单独起草,不宜直接翻译国内或 CE 文本——在第 8 条 §6 之下,一句多余的适应症足以改变整个项目的成本结构。其三,需逐件排查附件是否需单独定级,避免将一个巴西项目误算为单一申报。

Three practical points follow. First, classification belongs in the commercial decision phase, not the filing-preparation phase, because the class determines the three core costs: B-GMP, clinical evidence and timeline. Second, the intended-purpose statement should be drafted specifically for the Brazilian filing rather than translated from the domestic or CE text — under Article 8 §6, one superfluous indication is enough to change the cost structure of the whole project. Third, every accessory should be screened for whether it requires its own classification, so that a Brazilian project is not mistakenly scoped as a single filing.


本系列下一篇将就境外制造商的合规准入义务(持证主体的选任、授权书与 AFE/地方卫生许可的衍生义务)展开。

The next article in this series takes up the compliance-access obligations of foreign manufacturers — selecting the holder entity, and the derivative obligations attaching to the letter of authorization, the AFE and the local sanitary licence.


注释Notes

Notes

1. Lei nº 6.360, de 23 de setembro de 1976 —— 受卫生监督产品的注册与许可基本法。第 12 条:未经注册不得工业化、上架销售或交付;第 25 条:医疗器械(correlatos)制造或进口前须判定是否注册;第 2、50 条:仅经授权且场所已获地方卫生许可的企业方可从事生产、进口与储存。来源:Planalto, http://www.planalto.gov.br/ccivil_03/leis/l6360.htm

Lei nº 6.360 of 23 September 1976 — the framework statute on registration and licensing of products subject to sanitary surveillance. Art. 12: no product may be industrialized, offered for sale or delivered for consumption without registration; Art. 25: medical devices (correlatos) must be assessed, before manufacture or import, as to whether registration is required; Arts. 2 and 50: only authorized companies whose premises hold a local sanitary licence may manufacture, import or store such products. Source: Planalto, http://www.planalto.gov.br/ccivil_03/leis/l6360.htm

2. RDC nº 751, de 15 de setembro de 2022(现行合并文本,经 RDC 777/2023、RDC 810/2023 修订)—— 现行医疗器械核心规范,自 2023-03-01 生效。本文引用第 2 至 11 条、第 13、14 条及附件一。来源:DOU / ANVISA, https://anvisalegis.datalegis.net/action/ActionDatalegis.php?acao=abrirTextoAto&tipo=RDC&numeroAto=00000751&seqAto=000&valorAno=2022&orgao=RDC/DC/ANVISA/MS&cod_menu=1696&cod_modulo=134&pesquisa=true

RDC nº 751 of 15 September 2022 (current consolidated text, as amended by RDC 777/2023 and RDC 810/2023) — the core medical-device regulation, in force since 1 March 2023. This article cites Arts. 2 to 11, Arts. 13 and 14, and Annex I. Source: DOU / ANVISA (link above).

3. RDC 751/2022, Art. 2 —— 适用范围与排除(§2 二手/翻新→RDC 579/2021;§3 定制→RDC 305/2019;§4 IVD→条文援引 RDC 36/2015,现已由 RDC 830/2023 取代;§5 药品等;§6–§7 有源美学/清洁灭菌器械;§10–§11 附件豁免)。来源:同上。

RDC 751/2022, Art. 2 — scope and carve-outs (§2 used/reconditioned devices → RDC 579/2021; §3 custom-made devices → RDC 305/2019, repealed by RDC 925/2024; §4 IVDs → the text still refers to RDC 36/2015, since replaced by RDC 830/2023; §5 medicines etc.; §6–§7 active aesthetic and cleaning/sterilizing devices; §10–§11 accessory exemptions). Source: as above.

4. RDC nº 830, de 6 de dezembro de 2023 —— IVD 专项法规,采用“个人风险+公共健康风险”双重评估框架,设独立于本文所述二十二条规则的分类附件;其具体分类调整详见本系列文章一。来源:DOU / ANVISA, https://www.gov.br/anvisa/pt-br/assuntos/produtosparasaude/temas-em-destaque/arquivos/2024/rdc-830-2023-en.pdf

RDC nº 830 of 6 December 2023 — the dedicated IVD regulation, using a dual individual-risk/public-health-risk framework and its own classification annex, separate from the twenty-two rules discussed here; its classification changes are covered in Article One. Source: DOU / ANVISA (link above).

5.  RDC 751/2022, Art. 4 —— 医疗器械的实质性定义及“附件”(acessório,第 I 项)、短期/长期/瞬时使用(第 XLVI–XLVIII 项)等术语定义。来源:同注 2。

RDC 751/2022, Art. 4 — the substantive definition of a medical device and defined terms including accessory (acessório, item I) and short-term, long-term and transient use (items XLVI–XLVIII). Source: see note 2.

6.  RDC 751/2022, Arts. 6 & 7 —— 第 6 条:I/II 类适用 notificação(备案);第 7 条:III/IV 类适用 registro(注册)。来源:同注 2。

RDC 751/2022, Arts. 6 & 7 — Art. 6: Classes I/II are subject to notificação (notification); Art. 7: Classes III/IV are subject to registro (registration). Source: see note 2.

7. RDC 751/2022, Arts. 5、9 及 Anexo I —— 第 5 条:I 类低风险、II 类中风险、III 类高风险、IV 类最高风险,适用规则后仍存疑的由 ANVISA 定级(§2);第 9 条:依风险按附件一“规则 1–22”(Regras 1–22)分类。汇总表所列风险等级范围系根据各规则项下的分支情形整理,具体项目应以附件一葡文原文逐条核对。来源:同注 2。

RDC 751/2022, Arts. 5 and 9 and Annex I — Art. 5: Class I low, Class II medium, Class III high and Class IV maximum risk, with ANVISA determining the class where doubt remains (§2); Art. 9: classification by risk under Rules 1–22 of Annex I. The class ranges in the summary table are compiled from the branches of each rule; any specific product should be checked line by line against the Portuguese text of Annex I. Source: see note 2.

8. RDC 751/2022, Art. 8 —— 分类适用原则(§1 联合使用;§2 附件独立;§3–§4 软件;§5 最关键用途;§6 最严格规则优先;§7 连续使用时长的计算;§8 直接诊断)。来源:同注 2。

RDC 751/2022, Art. 8 — principles for applying the rules (§1 combined use; §2 accessories classified separately; §3–§4 software; §5 most critical use; §6 most stringent rule prevails; §7 computing continuous duration of use; §8 direct diagnosis). Source: see note 2.

9. RDC nº 657, de 2022 —— 软件医疗器械(SaMD)专门规定(标签、使用说明、安全与性能)。来源:DOU / ANVISA。

RDC nº 657 of 2022 — specific provisions on software as a medical device (SaMD), covering labelling, instructions for use, and safety and performance. Source: DOU / ANVISA.

10. RDC nº 305, de 2019(定制器械)与 RDC nº 579, de 2021(二手/翻新器械)—— 分别独立于 RDC 751/2022 的分类与注册路径。来源:DOU / ANVISA。

RDC nº 925 of 2024 (custom-made devices, replacing RDC nº 305 of 2019) and RDC nº 579 of 2021 (used/reconditioned devices) — each outside the RDC 751/2022 classification and registration pathway. Source: DOU / ANVISA.

11. RDC 751/2022, Art. 3 —— ANVISA 得对医疗器械家族、系统与成套(kit)授予备案或注册。来源:同注 2。

RDC 751/2022, Art. 3 — ANVISA may grant notification or registration to families, systems and sets (kits) of medical devices. Source: see note 2.

12. RDC nº 591, de 21 de dezembro de 2021(经 RDC nº 884, de 2024 修订)—— UDI 制度。表中所列为按风险等级分阶段的 UDI 赋码与标识附加义务截止日;可重复使用且载体附加于器械本体的产品顺延 2 年(第 15 条第 1 款)。SIUD 数据提交义务另依 IN nº 426/2026 分阶段起算,详见本系列文章一。来源:DOU / ANVISA, https://www.gov.br/anvisa/pt-br/assuntos/produtosparasaude/udi

RDC nº 591 of 21 December 2021 (as amended by RDC nº 884 of 2024) — the UDI regime. The table gives the class-phased deadlines for assigning UDIs and affixing the carriers; reusable devices bearing the carrier on the device itself receive a further two years (Art. 15 §1). SIUD data-submission obligations run on a separate phased timeline under IN nº 426/2026, covered in Article One. Source: DOU / ANVISA (link above).

13. RDC 751/2022, Art. 10(并参第 13、14 条的文件清单)—— §1 注册文档审评;§3 补充文件要求;§4 不发技术补正要求、径行驳回的事由;§5 备案不作技术审查;§6 备案 30 日内处理;§7 维持须符 B-GMP 与基本要求;§8 注册以 B-GMP 证书公布为条件;§9–§10 语言要求。来源:同注 2。

RDC 751/2022, Art. 10 (read with the document lists in Arts. 13 and 14) — §1 review of registration documentation; §3 requests for additional documents; §4 grounds for summary rejection without a technical deficiency letter; §5 no technical analysis of notifications; §6 notifications processed within 30 days; §7 maintenance conditional on B-GMP and essential requirements; §8 grant of registration conditional on publication of the B-GMP certificate; §9–§10 language requirements. Source: see note 2.

14. RDC 751/2022, Art. 11 —— 注册自 DOU 公布起有效期 10 年,可以相同期限连续再验证(revalidação)。来源:同注 2。

RDC 751/2022, Art. 11 — registration is valid for 10 years from publication in the DOU and may be revalidated successively for equal periods. Source: see note 2.

15. RDC nº 848, de 6 de março de 2024(DOU 公布 2024-03-08,生效 2024-09-04)—— 医疗器械与 IVD 基本安全与性能要求,取代 RDC 546/2021;引用其关于 III/IV 类临床数据要求的条款。来源:DOU / ANVISA, https://anvisalegis.datalegis.net/action/ActionDatalegis.php?acao=abrirTextoAto&tipo=RDC&numeroAto=00000848&seqAto=000&valorAno=2024&orgao=RDC/DC/ANVISA/MS&cod_menu=9434&cod_modulo=310&pesquisa=true

RDC nº 848 of 6 March 2024 (published in the DOU on 8 March 2024, in force from 4 September 2024) — essential safety and performance requirements for medical devices and IVDs, replacing RDC 546/2021; Arts. 15–16, 18 and 19 govern clinical evidence generally and device-specific clinical investigations, including the Class III/IV cases in Art. 19, item II. Source: DOU / ANVISA (link above).

16. Lei nº 6.437, de 20 de agosto de 1977 —— “界定违反联邦卫生立法之行为并设定相应制裁”;未注册/未备案即行销售构成卫生违法,可致罚款、扣押销毁与暂停商业化。来源:Planalto, http://www.planalto.gov.br/ccivil_03/leis/l6437.htm

Lei nº 6.437 of 20 August 1977 — defines infringements of federal sanitary legislation and the corresponding sanctions; selling without registration or notification is a sanitary infraction that may lead to fines, seizure and destruction, and suspension of commercialization. Source: Planalto (link above).